- Process noncritical patient-care devices using a disinfectant and the concentration of germicide listed in Table 1. Category IB.17, 46-48, 50-52, 67, 68, 378, 382, 401
- Disinfect noncritical medical devices (e.g., blood pressure cuff) with an EPA-registered hospital disinfectant using the label’s safety precautions and use directions. Most EPA-registered hospital disinfectants have a label contact time of 10 minutes. However, multiple scientific studies have demonstrated the efficacy of hospital disinfectants against pathogens with a contact time of at least 1 minute. By law, all applicable label instructions on EPA-registered products must be followed. If the user selects exposure conditions that differ from those on the EPA-registered product label, the user assumes liability from any injuries resulting from off-label use and is potentially subject to enforcement action under FIFRA. Category IB. 17, 47, 48, 50, 51, 53-57, 59, 60, 62-64, 355, 378, 382
- Ensure that, at a minimum, noncritical patient-care devices are disinfected when visibly soiled and on a regular basis (such as after use on each patient or once daily or once weekly). Category II. 378, 380, 1008
- If dedicated, disposable devices are not available, disinfect noncritical patient-care equipment after using it on a patient who is on contact precautions before using this equipment on another patient. Category IB. 47, 67, 391, 1009